Kinetic Image Cytometer System Power Up Procedure
The Kinetic Image Cytometer system is designed such that all components can be left 'on' indefinitely. However, during any major system updates, rare system crashes, power failures etc, the following sequence should be used.
Turn on the Electrical Stimulator module.

The Electrical Stimulator module should turn on the PC/Workstation automatically. If not, manually turn on the Workstation. Turn on the Workstation Monitor.

Turn on the system Control Unit module. Make sure the STAGE CONTROL switch is in ENABLE position.

Turn on the Camera Control module.

Turn on the Light Source module (located at the back of the system enclosure).

Turn on the Microscope module (located on the right side of the system enclosure).